Hanna HI 9813-6 - Waterproof pH/EC/TDS/Temperature Meter with pH Electrode Diagnostic

Hanna HI 9813-6 - Waterproof pH/EC/TDS/Temperature Meter with pH Electrode Diagnostic

Description

pH electrodes are commonly designed to produce and output signal equal to 0 mV when immersed in a solution with pH 7. But a constant output value at pH 7 occurs as a result of certain adverse events, such as:

A broken bulb in the pH electrode
A short circuit in the cable or connector
Magnetic force leakage in the cable or connector (for example, in case of excessive humidity)

In these situations, the instrument displays a false reading at pH 7 and the user may not be aware that corrective measures are necessary; particularly if the user is anticipating a pH neutral reading. To avoid this problem, HANNA developed F.S.T. technology for HI 9813-6.

With this special security system, a system failure will force a fixed value close to pH 4, rather than pH 7, and render the problem clearer to the user.

For added confidence in your measurements, the meter is equipped with two LEDs and a "CHECK" button. While pressing this button, upon immersing the electrode in the Cal-Check solution HI 50021P, one of the LEDs will light up to indicate the status of the meter. The green LED ("OK") indicates the probe is working correctly, while the red LED ("C&C" Clean & Cal) indicates the electrode needs attention.

HI 9813-6 is supplied with pH/EC/TDS/T HI 1285-6 probe with 1 m (3.3') cable, pH 7, 1.41 mS/cm and 1500 ppm calibration solution (20 mL each), cleaning solution (2 pcs), HI 50021 test solution (2 pcs), blue protective case, battery and instructions.

Range pH 0.0 to 14.0 pH
Range EC 0.00 to 4.00 mS/cm
Range TDS 0 to 1999 ppm (mg/L)
Range Temperature 0.0 to 60.0C
Resolution pH 0.1 pH
Resolution EC 0.01 mS/cm
Resolution TDS 1 ppm (mg/L)
Resolution Temperature 0.1C
Accuracy pH 0.1 pH
Accuracy EC 2% F.S.
Accuracy TDS 2% F.S.
Accuracy Temperature 0.5C
TDS Conversion Factor 0.56 to 0.72 ppm = 1 S/cm
Calibration manual at 1 point (all parameters except temperature)
Temperature Compensation automatic, 0 to 50C (32 to 122F) with = 2%/C (EC/TDS only)
Probe HI 1285-6 pH/EC/TDS/T,
with Clean & Cal, 1 m (3.3') cable (included)
Battery Type / Life 1 x 9V / approx. 150 hours of continuous use
Environment 0 to 50C (32 to 122F); RH max 100%
Dimensions 165 x 75 x 45 mm (6.9 x 2.9 x1.8")
Weight 180 g (6.3 oz.)

Item Number: 131895

Documents

pdf file Literature
pdf file Instruction Manual

This item is not available to order. Browse our items available to order.


Related Items

Home | Products | Learning Center | About Us | Returns & Refunds | Shipping | Privacy | Testimonials | Contact | Login | View Cart |

Copyright 1992-2019 Greentrees Hydroponics
2581 Pioneer Avenue Suite D, Vista, California 92081